Calibration – Radon Stripping (Test-Lines) 

Radon Stripping (test-lines) calibration aliases and parameters control file

//_____________________________________________________________________________

//

//  ALIASES


    spectraDownName     spectrum_down   // downward-looking spectra

    spectraUpName       spectrum_up     // upward-looking spectra

    livetimeName        liveTime        // live time (secs)



//  PARAMETERS


    includeUpward              1        // 0 – dont include upward-looking calibration

                                        // 1 – include upward-looking calibration


    useSumSpecEncal            1        // 0 – do encal on individual lines 

                                        // 1 – use sum spec of all lines for encal


    encalMethod                2        // 1-automatic (cross-correlation)  2-manual(peak identification)


    thoriumPeakChannel       205        // channel number at thorium photopeak maximum


    calibrationPeakEnergy  0.609        // calibration peak energy (MeV)

    calibrationPeakEnergy  1.765        // calibration peak energy (MeV)


//_____________________________________________________________________________